Ponsel Turbidity Sensor

Ponsel leads the way in turbidity measurement with its innovative probes. All sensors are optical, with advanced technology for accurate and reliable turbidity measurements. They cover a wide range of applications in both low and high levels, such as drinking water, wastewater, surface, or process water. With Modbus RS-485 and SDI-12 digital communication, our probes offer an easy ready-to-use solution to integrate turbidity into an existing infrastructure.

Technical specifications

Measures

Measure principle Diffusion IR at 90°
Measure ranges 5 to 4000 NTU in 5 ranges:

• 5 – 50 NTU

• 5 – 200 NTU

• 5 – 1000 NTU

• 5 – 4000 NTU

• AUTOMATIC

0 to 4500 mg/L

• Calibration : Range 0-500 mg/L according to NF EN 872

• Range >500 mg/L according to NF T 90 105 2

Resolution 0,01 to 1 NTU – mg/L
Accuracy < 5% of the reading
Working temperature 0°C to + 50°C
Measure of temperature Via CTN
Stocking temperature -10°C to + 60°C
Signal interface Modbus RS-485 (standard) and SDI-12 (option)
Maximum refreshing time < 1 second
Sensor power-supply 5 to 12 volts
Electric consumption Standby: 40 μA
Average RS485 (1 measure/seconde): 820 μA
Average SDI12 (1 measure/seconde): 4,2 mA
Current pulse: 500 mA

Sensor

Dimensions Diameter : 27 mm; length : 170 mm
Weight 300 g (sensor + cable 3 meters)
Material PVC, DELRIN, Quartz, PMMA, Polyamide
Maximum pressure 5 bars
Connection 9 armoured connectors, polyurethane jacket, bare-wires or waterproof Fisher connector
Degree of protection IP68